= SESANS Session Notes SAS2024 - Sun 2024-11-03 =
 
=== Topic: Cross-calibration ===
* Following round-robin reference samples have been used so far for SESANS at several sources (according to Wim Bouwman): Flexi-Por, Papyex. These measurements have also been compared to SANS/USANS results from the same samples.
* But: these kinds of samples have the disadvantage that they show artifacts when used for TOF-SESANS.
* Also, the presented reference data had differences in peak-minima of measured reverence samples at different sources. The reason for this difference still has to be identified.
* General consens: We need more sets of standard samples at all facilities
 
=== Notes from the discussion round: ===
 
* What is needed: New standard samples with shorter length scales of around 50nm (but such gratings are difficult to make)
* The question was raised if new round robin samples should be fabricated which can be handed from facility to facility. On the other hand, a good reference sample depends on each use-case - there is no perfect standard sample for several techniques/facilities. Instead, good existing standard samples (like the above mentioned and already measured Flexi-Por/Papyex) should be used.

Latest revision as of 13:58, 12 March 2025

Notes from canSAS 2024 Second Working Group Discussion Session

Includes:

  • Data Formats
  • TOF SANS
  • SESANS
  • Tomograph

SESANS Session Notes SAS2024 - Sun 2024-11-03

Topic: Cross-calibration

  • Following round-robin reference samples have been used so far for SESANS at several sources (according to Wim Bouwman): Flexi-Por, Papyex. These measurements have also been compared to SANS/USANS results from the same samples.
  • But: these kinds of samples have the disadvantage that they show artifacts when used for TOF-SESANS.
  • Also, the presented reference data had differences in peak-minima of measured reverence samples at different sources. The reason for this difference still has to be identified.
  • General consens: We need more sets of standard samples at all facilities

Notes from the discussion round:

  • What is needed: New standard samples with shorter length scales of around 50nm (but such gratings are difficult to make)
  • The question was raised if new round robin samples should be fabricated which can be handed from facility to facility. On the other hand, a good reference sample depends on each use-case - there is no perfect standard sample for several techniques/facilities. Instead, good existing standard samples (like the above mentioned and already measured Flexi-Por/Papyex) should be used.